技术中心

您现在的位置:化工生意网 > 技术首页 > 技术交流

On-Site Mobile NH3 Measurements at Livestock Farm

2026年08月04日 09:48:18人气:28来源:


On-Site Mobile NH3 Measurements at Livestock Farm       


       Here we show a typical ammonia mobile analyzing system in Fig. 1. It contains the HealthyPhoton HT8700 ammonia analyzer, Campell’s CR3000 data acquisition module, GPS module, ultrasonic 3D anemometer module, and real-time data processing module. The whole system can be installed on the automobile roof rack and collect 10Hz high-speed ammonia concentration data in real-time. The latitude, longitude location data, wind direction, and wind speed data are collected simultaneously in real-time by GPS module and anemometer module. All the data are synchronized and transmitted to cloud for remote data processing and analysis to get the real-time ammonia concentration, providing a complete solution for high speed and accurate ammonia emission monitoring and tracing.


Mobile NH3 sensing platform at a farmland based on HT8700

Fig. 1 Mobile NH3 sensing platform at a farmland  based on HT8700


       Fig. 2 shows the tested atmospheric ammonia concentration surrounding a pig farm at Ningbo. The red pin indicates the location of the farm, and the color trace is the testing route. Blue means low concentration while red means high. We found two peaks of concentration (>250 ppb) in the experiment, which were at the main entrance of the farm (349.3 ppb) and the manure dumping area, respectively. They were both ammonia emission sources, and the concentration decreased gradually as the distance increases. Fig. 3 is a sample of the transient data acquired from the experiment. It is shown that HT8700 responded to the concentration changes very quickly (response time ~ 1s). The experiment proved the system stability and reliability on a vehicle traveling for hundreds of kilometers while achieving high-speed and high-accuracy measurements of sub-ppb level ammonia.


Measured NH3 concentration map around a pig farm

Fig. 2 Measured NH3 concentration map around a pig farm


HT8700 ammonia analyzer showing fast responses to concentration changes

Fig. 3 HT8700 ammonia analyzer showing fast responses to concentration changes

全年征稿/资讯合作 联系邮箱:31food@vip.qq.com
版权与免责声明
1、凡本网注明"来源:化工生意网"的所有作品,版权均属于化工生意网,转载请必须注明化工生意网,https://www.chem31.com。违反者本网将追究相关法律责任。
2、企业发布的公司新闻、技术文章、资料下载等内容,如涉及侵权、违规遭投诉的,一律由发布企业自行承担责任,本网有权删除内容并追溯责任。
3、本网转载并注明自其它来源的作品,目的在于传递更多信息,并不代表本网赞同其观点或证实其内容的真实性,不承担此类作品侵权行为的直接责任及连带责任。其他媒体、网站或个人从本网转载时,必须保留本网注明的作品来源,并自负版权等法律责任。
4、如涉及作品内容、版权等问题,请在作品发表之日起一周内与本网联系,否则视为放弃相关权利。